|
Analytical Systems
Introduction
Direct-push
analytical systems are attachments designed to be used with direct-push platforms.
These systems include a diverse and growing class of instruments that are adapted
for in situ use as part of the direct-push tooling. Direct-push analytical instrumentation
allows real-time or near real-time data to be generated in the field while sampling,
without the many requirements associated with sample management and while generating
minimal investigation-derived waste. This "all-in-one" approach potentially
allows the user to conduct a more rapid and detailed assessment at a lower overall
cost than they could achieve with more traditional methods such as drill rigs
and off-site laboratories.
Description
Some of the systems incorporate a sensor such as a laser-induced
fluorescence (LIF) or x-ray fluorescence
(XRF) directly into the probe that is advanced with the direct-push tooling
into the subsurface. Others are sophisticated closed systems that retrieve volatile
organic compounds (VOC) from the subsurface and route them into an integrated
instrument for analysis. What all of these systems have in common is that they
allow the user to quickly characterize a site in the field using relatively
agile and minimally intrusive direct-push platforms.
Typical Uses
Direct-push
analytical systems can be used for practically any environmental purpose, including
site assessment, site characterization, removal assessment, or even monitoring
of natural attenuation. Practical uses for these systems are partly a function
of the platforms themselves, in situ field analyses, and specifications of the
particular system. General applications of direct-push platforms and in situ
field analysis are described in more detail below; the characteristics and applications
of several classes of direct-push analytical system are described in the following
sections.
As described in the entry of the same name, direct-push
platforms are highly versatile and efficient means of accessing and sampling
sites with shallow subsurface contamination. Briefly, these platforms, which
include both rotary hammer and cone penetrometer (CPT) rigs, advance samplers
and analytical instrumentation into the subsurface by hydraulically pushing
a string of rods into the ground. Rotary hammer units are deployed on a truck
or van, although units have been developed for a wide variety of vehicles including
small four-wheel all-terrain vehicles and several sizes of tracked vehicles.
Rotary hammer units use a hydraulic ram with a rotary hammer, combining the
static force of the platform and the percussive force of the hammer to advance
the tool string into the subsurface. CPT rigs are more massive than rotary hammer
units, typically weighing anywhere from 15 to 30 tons. A hydraulic ram that
uses the static force of the vehicle alone is sufficient to advance the tooling
into the ground for this reason. Both units can advance a variety of soil and
soil gas samplers, groundwater samplers, geotechnical sensors, and analytical
instrumentation into the subsurface. Many of the sampling and analytical systems
originally developed exclusively for one platform or the other are being adapted
for both, creating a great deal of cross-pollination between the two and allowing
users a growing range of choices for using these direct-push platforms.
The very nature
of the direct-push platforms leads to certain types of uses for direct-push
analytical systems. By virtue of their size, weight, and limited height these
platforms are agile in comparison to a drill rig and can be used to access hard-to-reach
areas. They are often used inside buildings or in tight outdoor spots with limited
overhead clearance. Little waste is produced. No drill cuttings are generated
because the tooling is pushed into the ground. Conversely, direct-push platforms
can only advance a sampler or instrument into unconsolidated soils and sediments
and are limited in the depths they can achieve (between 60 and at most about
100 feet for a rotary hammer, and about 100 to 150 feet for a CPT under ideal
conditions). For these reasons, these analytical systems are typically used
to assess and characterize sites with shallow soil and groundwater contamination,
and often are used on sensitive or hard to reach sites such as residential property
or inside industrial buildings.
In situ field screening or analysis of any kind is limited in the precision
and accuracy of the results that can be generated. Samples are analyzed directly
in the subsurface or are brought to the surface to be analyzed. While there
may be some inherent advantages to analyzing samples in situ (such as a reduced
loss of VOCs or elimination of human error in sample handling), precision and
accuracy of the sample results will vary from formal laboratory results because
of the lack of controls on sample collection and analysis. For example, soils
and sediments may not be in direct contact with the sensor window, sample volumes
may be inexact when compared with a precision balance used in the laboratory,
and sample collection may be affected by subsurface conditions such as extremely
tight clays. Some direct-push analytical sensors such as LIF are not compound-specific
but provide a measure of total contaminants in a class.
Direct-push analysis is well suited for application in a
triad
approach to conducting site characterization and removal monitoring, particularly
voluntary cleanup and Brownfield sites. The triad approach uses on-site analytical
tools in conjunction with systematic planning and dynamic work plans to streamline
sampling, analysis, and data management conducted during site assessment, characterization,
and cleanup. Field analysis in general and direct-push systems in particular
are often used to speed collection and reduce costs on projects where the sites
are large, a high volume of data points are needed, the sites are partly or
totally inaccessible by a large drill rig, or to minimize sampling disturbances
in sensitive habitats. Despite the inherent limitations of field analysis, careful
use of direct-push platforms and analytical systems can reduce the overall uncertainty
surrounding site contamination.
A variety of analytical instrumentation systems are presented in detail in the
following sections. These sections provide information on systems for both organic
and inorganic screening and analysis, including a general description and typical
uses of the systems, the basic theory of operation, components, and information
on target analytes and performance. The general advantages and limitations of
the systems are also provided, along with links to the manufacturers and vendors.
A table summarizing
and comparing the available technologies may be viewed by clicking here.
Fluorescence
Instrumentation
DESCRIPTION
Fluorescence instrumentation
refers to a general class of technologies in which a subsurface sample is bombarded
with an ultraviolet (UV) light source that causes petroleum hydrocarbons in the
sample to fluoresce. The resulting fluorescence is used to identify and measure
the contaminant. The fluorescence is produced when molecules of a certain class
of VOCs, known as polycyclic aromatic hydrocarbons (PAH), give off energy in an
attempt to return to their "natural" or ground state after becoming
excited by the energy from the UV source.
Fluorescence instruments were originally developed for use by CPT rigs but are
now also deployed from rotary hammer rigs. Several energy sources, including lasers
and mercury lamps, have been used in these instruments. This section will discuss
the mercury (Hg) lamp fluorescence detector. The fiber-optic based LIF chemical
detection system allows for real-time, in situ subsurface detection of fuel hydrocarbon
contaminants. An encyclopedia devoted entirely to LIF may be viewed by following
this link:
Laser Induced Fluorescence
Figure 1.
 |
|
Source:
Applied Research Associates, Inc. 1999
|
TYPICAL USES
Like LIF, Hg fluorescence
instruments are used to rapidly characterize a site accessible to direct-push
platforms. With direct-push equipment, Hg fluorescence detectors (Figure 1)
cut the time needed to delineate the extent of hydrocarbon plumes from fuel
spills or leaking storage tanks. Hg instruments are used for site-specific relative
screening of contamination levels.
THEORY OF OPERATION
As
with LIF, a subsurface sample is bombarded with UV light from the Hg source,
causing certain petroleum hydrocarbons to fluoresce. The sample absorbs the
source energy, which results in the elevation of electrons from orbitals in
the ground state to higher energy orbitals in an excited state. When the excited
electrons return to the ground state, light energy is released as fluorescence
emission spectra that can be measured and analyzed (Silverstein and others 1991).
The absorption of UV source energy is dependent on the electronic structure
of the organic molecule. Compounds consisting of double carbon bonds with weakly
attached electrons (such as PAHs) can be identified using low-energy bombardment
techniques. The source energy dictates which specific analytes and compounds
can be detected (Keller and others 2001).
The resulting fluorescence
is measured using an electronic detector and the reading is converted into a
contaminant concentration.
SYSTEM COMPONENTS
A common commercial Hg system consists of a down-hole module and an up-hole
controller. The down-hole module can be attached directly to CPT geotechnical
instruments to provide simultaneous lithologic analysis and contaminant screening.
A 254-nanometer Hg lamp is used to project UV light through a sapphire window
in the side of the probe to excite the sample. The resulting fluorescence is
returned as an optic signal to the controller module on the surface via a cable
that extends inside the direct-push tool string. Unlike LIF, the signal light
is converted to an electric current for amplification and signal conditioning,
which reduces the hardware cost of the system and improves its ease of operation.
Using the controller, the operator is able to adjust for background fluorescence
levels in soils with naturally occurring fluorescent substances.
MODE OF OPERATION
When the Hg fluorescence
instrument is used with CPT equipment, it provides a continuous output of fluorescence
over the entire depth of the investigation. This information can be viewed graphically
in real time using a data acquisition system as the probe is advanced, providing
an easily interpreted view of the plume (Figure 2). The continuity of the data
reduces the time and effort required for data interpretation and presentation.
Multiple profiles across a site can be used to develop a three-dimensional model
of the plume during the site characterization, without the time or expense of
off-site sample analysis or post-processing of the data.
 |
| Source: Applied
Research Associates, Inc. 1999 |
TARGET ANALYTES
An Hg system can
be configured to detect a wide range of contaminants such as jet fuel, diesel,
unleaded gasoline, home heating and motor oil, and by applying filters, coal
tars and creosote.
PERFORMANCE SPECS
Interferences
Although intended to specifically target petroleum hydrocarbons, the excitation
energy produced by the UV source may cause other substances to fluoresce as
well, which may cause interference problems. Many commonly occurring fluorescent
minerals such as calcite can produce a measurable signal. Other man-made non-hydrocarbon
fluorescent material may be found in the subsurface environment, such as deicing
agents and antifreeze additives. Many detergent products are known to fluoresce
very strongly. Naturally occurring organic matter, which includes PAHs, also
can fluoresce. An experienced operator may be able to differentiate between
the fluorescent signatures of hydrocarbons and other interfering compounds.
Environmental conditions, including temperature and moisture may affect the
performance of the instrumentation. The output of the Hg lamp is reduced in
cold temperatures, resulting in lower sensitivity. It is recommended that the
probe be stored in a warm place prior to testing. Moisture effects on sample
results are measurable but are generally not significant.
Detection Limits
Detection limits are soil and fuel dependent. Total petroleum hydrocarbon values
as low as 100 parts per million (ppm) can be detected in sandy soils. Detection
limits may be reduced in clays because clays have a greater surface area than
sandy soils. More fuel is locked within the clay structure, obscuring a portion
of the signal that is received by the detector. In addition, fuels with a lower
proportion of PAHs may have higher detection limits, and additives such as dyes
(sometimes used for taxation purposes) may obscure the response and raise detection
limits.
Calibration
The Hg fluorescence instrumentation is field calibrated before and after
each test. Calibration is relatively simple. A special card that has a black
surface and a gray surface is placed in front of the probe. The black surface
gives no response to the detector, providing a baseline “zero” reading for the
instrumentation. The gray surface gives a response that is about 50 percent
of a full-scale output.
QUALITY CONTROL
Quality control
for down-hole analytical instruments consists of proper instrument calibration,
and confirmatory samples.
The field calibration
method is repeated after sampling as a calibration check, to determine whether
any shift occurred in the detector during use. If a shift is detected in the
instrument, the pre- and post-calibration results can be used to correct the
data during post-processing. If the shift is very large, the instrument may
have to be returned to the factory for evaluation. Confirmatory samples are
collected by advancing samplers adjacent to the down-hole sample location. These
samples are collected from intervals exhibiting varying levels of contamination
(including non-detect) and sent to a formal laboratory for analysis. The analytical
results are then compared to the in situ results to assess the technology's
performance across the entire range of contaminant concentrations.
Precision and Accuracy
In general, fluorescence
units are semi-quantitative and semi-qualitative in nature. Only the PAHs in
fuel fluoresce when bombarded by the UV source; different fuel fractions are
qualitatively identified by analyzing the wavelength at which they fluoresce
to determine which PAHs are present in what ratios, and this information is
compared to standards for various fuels. The correlation between fluorescence
and hydrocarbon contamination in a given soil matrix is generally linear as
represented in Figure 3. This relationship is used to convert the site-specific
fluorescence levels into relative contamination concentration levels.
 |
| Source: Applied
Research Associates, Inc. 1999 |
Membrane
Interface Probe
DESCRIPTION
| Figure 4. |
 |
The
permeable membrane interface probe (MIP) was developed to allow for near real-time
evaluation of subsurface VOCs using rotary hammer units. The MIP is now used
widely with CPT rigs as well. The MIP probe consists of a thin composite metal
and a Teflon membrane impregnated into a stainless steel screen on the face
of a probe. The probe is mounted on a standard direct-push rod (see Figure 4).
A carrier gas line runs from the probe to the detector through the inside of
the tooling, and can be connected to several types of detectors, including flame
ionization detectors (FID), photoionization detectors (PID), or direct sampling ion trap
mass spectrometers (DSITMS). The device allows the user to detect VOCs as
it is driven to depth. VOCs are drawn through the system's semi-permeable membrane
and carried to a detector at the surface where they are analyzed and measured.
The MIP also incorporates a lithologic sensor and logs advancement speed. Use
of the MIP allows investigators to identify the contaminant distribution and
migration pathways in real time.
TYPICAL USES
The MIP system
may be used to characterize any site with shallow subsurface VOC contamination,
including sites with fuel releases, chlorinated solvent releases, and dense
non-aqueous phase liquid (DNAPL). The technology is only applicable to shallow
contamination; standard components include a 60-foot gas carrier line, but with
a custom-built gas carrier line the practical vertical limit of characterization
is about 100 feet when deployed with a rotary hammer, and could be as deep as
150 feet if driven with a CPT rig. The system is used to simultaneously characterize
the subsurface soils and sediments with chemical contamination. Using the MIP,
electrical conductivity logs, and advancement speed logs, it provides information
on contaminant distribution and migration pathways.
The MIP is typically used with a PID or FID detector (or both in series) to
rapidly delineate the distribution of VOCs in the subsurface. These detectors
have the advantage of providing rapid, continuous measurements of relative VOC
concentrations in the subsurface. The data collected with the MIP can be used
to accurately place a minimum number of conventional sampling points (soil bores
and monitoring wells) for site characterization and monitoring.
Figure 5.
 |
| Source: U.S.
Army Environmental Center 2000 |
The DSITMS (Figure
5) may be used instead of the PID or FID as a high-level field screening technique
to provide qualitative identification of specific compounds. The DSITMS is a
versatile tool for fast on-site qualitative or quantitative measurement of organic
compounds in air, water, soil, and wastes. The DSITMS is best suited for analyzing
samples for the presence of VOCs in support of site activities requiring the
analysis of large numbers of samples in a short period of time, or for routine
quantitative monitoring of sampling locations that have been previously characterized.
THEORY OF OPERATION
In practice, the
MIP membrane is heated to between 80° C and 125° C as it is advanced
through the subsurface. VOCs present in the subsurface partition into the membrane
and migrate through it by diffusion, rather than being pulled in by a vacuum
or actively purged using an inert gas, as with the Hydrosparge system (described
below). The VOCs move across the membrane into a helium carrier gas that flushes
the back of the membrane and transports the VOCs to the aboveground detector.
Unlike many analytical instruments, the system can operate in both the vadose
zone and beneath the water table.
Relative concentrations
of aromatic VOCs may be measured continuously using a PID during advancement
of the probe, while a FID is used to detect less volatile, straight chain hydrocarbons;
both detectors may be used in series. The PID and FID detectors are not compound-specific
but measure the total response of all volatile compounds in the sample that
can be ionized by the detectors and are present in sufficient concentrations
to be detected by the detector. The PID and FID units are not quantitative;
the electrical response of the detector to VOCs in the subsurface is registered
in micro-volts (µV). The range of response to similar contaminants varies
from site to site. For example, the detector response for gasoline-range organics
has been found to vary from 4,000 to 50,000 µV per milligram per kilogram
(mg/Kg) in soil, depending on the soil type (Geoprobe Systems 1996). For this
reason, the relative responses of the detector may be used as a general indication
of the concentration of VOCs present in the subsurface.
For qualitative identification of specific compounds, a DSITMS is used. The
MIP is capable of multiple, discrete VOC measurements in a single penetration,
and is a versatile tool for a fast on-site qualitative or quantitative measurement
of organic compounds in air, water, and soil. Much progress has been made in
recent years developing and validating methods for the screening and quantification
of targeted VOCs in discrete water and soil samples. Detection limits are typically
in the range of 1 ppm or less with little or no sample preparation required,
and a sample analysis time of 2 to 3 minutes. The DSITMS field methodology (draft
Method 8265) has been conditionally approved by the U.S. Environmental Protection
Agency (EPA).
Subsurface
lithology is determined by comparing readings from the integrated electrical
conductivity sensor with the rate of advance through the subsurface, in much
the same way as CPT sensors compare sleeve friction to tip resistance readings.
Briefly, an electrical current is passed between two opposing poles of the conductivity
sensor, and the conductivity of the soils is measured as the probe is advanced.
Different classes of soils have different specific conductance, and this property
is used to identify different soil layers during advancement. The speed of penetration
is also logged and can be compared with conductivity to confirm the results,
as the rate of advance differs by soil class.
SYSTEM COMPONENTS
The MIP system
consists of the probe, a "trunkline" containing the power cable, data
cable, and gas line, an electronic controller and data output system. An electrical
conductivity sensor is also coupled with the MIP to provide a real-time log
of subsurface soils based upon changes in their electrical resistivity. Components
of this system include the sensor and data acquisition "stringpot"
cable. The cabling runs through the inside of the tool string as well. An advancement
speed gauge is attached to the hammer probe when used with a rotary hammer platform.
MODE OF OPERATION
As the operator
advances the MIP sensor into the subsurface, a log is displayed on the field
computer's screen. This log provides information about total VOC contamination
obtained from the detector(s). The real-time log (see Figure 6) also provides
a depth and speed graph, conductivity log of the soils, and temperature log
of the heated sensor.
Figure 6.
MIP Data Log

|
| Source: Kejr, Incorporated 2001 |
Running MIP logs
on a grid or targeted pattern across an investigation area will provide a three-dimensional
view of VOC distribution and lithology. Optional software allows construction
of cross sections from the MIP and conductivity logs (see Figure 7).
 |
| Figure 7.
Source: Kejr, Incorporated 2001 |
Information from
both the MIP and electrical resistivity logs provides information on contaminant
distribution and migration pathways. This information enhances development of
an accurate site conceptual model, which can significantly reduce the cost of
investigation activities.
TARGET ANALYTES
An assortment of
VOCs in the vadose zone and in groundwater may be measured and possibly identified,
depending on the detector used with the system.
PERFORMANCE SPECS
Interferences
The PID and FID
units are total VOC detectors. Because the MIP system does not incorporate a
gas chromatograph (GC) for compound separation prior to analysis, the system
cannot differentiate VOCs if these detectors are used. Any VOCs that are not
targets of the investigation may interfere with the instruments' response. For
example, it would not be possible to delineate a jet fuel release if diesel
or gasoline had also been released into the subsurface.
The DSITMS does
not use a GC for compound separation. A series of scans containing ions indicating
the presence of VOC analytes is used to qualitatively identify analytes of interest.
Multiple scans of standards containing the target analytes are integrated, calibrating
the instrument to identify these compounds; however, ions with a similar mass
and charge to the target analytes' characteristic ions could cause interference
and artificially increase the reported concentrations.
A blank sample
should be run for background subtraction, to ensure there is no carryover in
the transfer line, any time samples are run having greater than 500 ppm of VOC
contamination, and between analysis of samples from different sources. A system
blank check should be performed before and after each set of in situ measurements
when using the DSITMS detector.
Detection Limits
The detection limit
for typical chlorinated compounds using the PID and FID detectors is about 5
ppm. The DSITMS detector can detect hydrocarbons as low as 1 ppm in solution,
depending on the subsurface conditions.
Calibration
Calibration of
the MIP system when equipped with the PID and FID detectors consists of response
testing. Response testing demonstrates to the operator that the system is working
and is generating a response to environmental contaminants. A standard containing
a known quantity of fuel related hydrocarbons or chlorinated solvents is prepared
and mixed into 0.5 liter of fresh water. The MIP probe is first immersed into
a bucket of clean sand saturated with fresh water, in order to stabilize the
baseline response of the detector to water (the PID is by nature sensitive to
moisture on the ultraviolet lamp used to ionize target compounds). The probe
is then placed into the water containing the standard for 45 seconds, and the
response is observed.
Calibration of the DSITMS is run initially at project start up. A calibration curve
is developed using laboratory-prepared standards of known concentrations bracketing
the expected contaminant concentrations for a particular site. Critical operating
parameters used to generate the calibration curve, including carrier gas flow
rate of the MIP, membrane temperature, and DSITMS settings, should not be changed
during sample analysis.
Quality Control
For the PID and
FID detectors, response testing should be performed before each MIP log.
Externally-prepared calibration check standards should be run at the startup
and at the end of each day of operation using the DSITMS to confirm the unit's
calibration. An acceptable calibration compound such as perflourotributylamine
should be used.
A minimum of 5 percent verification sampling and off-site analysis by EPA
Method 8260B should be performed to confirm the results obtained at a particular
site using the DSITMS. Locations and depths for verification sampling should
be selected to include a range of contaminant concentrations from non-detect
to the maximum concentrations detected by the DSITMS. PID and FID readings may
also be compared to confirmatory samples such as these to provide a general
indication of the concentrations observed at other locations during operations.
Precision and Accuracy
The PID and FID
detectors do not provide quantitative results. The relative response of the
PID and FID can vary as much as an order of magnitude depending on site conditions
and soils.
The DSITMS has demonstrated the capability of meeting the precision and accuracy
quality control (QC) performance criteria established for water analysis by
Continuing Calibration Check - EPA Method 624 (40 Code of Federal Regulations
[CFR] Part 136). Validation data collected from saturated soils and compared
to data for samples collected by EPA
Sampling Method 5035 and Analysis
Method 8260B indicate that the system provides quantitative estimates of
subsurface contamination distribution. Quantitative results appear to be less
precise for vadose soils
Hydrosparge
DESCRIPTION
 |
Figure 8.
Source: U.S. Army Environmental Center 2000 |
The Hydrosparge
system shown in Figure 8 is similar to the MIP system in that it extracts VOCs
from groundwater and brings them to the surface for analysis via a closed system.
It differs from the MIP in a couple of respects. Unlike the MIP, the Hydrosparge
is active and physically purges VOCs from the sample interval rather than allowing
them to passively diffuse into the sampler. The Hydrosparge is only able to
sample one discrete interval, as the probe must be retracted to expose the sampler
and cannot be re-advanced (multiple depths may be characterized within the sampler
screen). The Hydrosparge does not incorporate a lithologic sensor of any sort.
Sampling intervals are selected based upon separate pushes with CPT sensors
at the desired sample location. Standard CPT sensors are deployed in a probe
that measures the resistance against the cone tip and friction against the cone
sleeve; system software estimates the makeup of subsurface soils and sediments
by measuring and comparing these mechanical forces.
The
Hydrosparge system is designed to collect VOCs from groundwater for real-time
analysis by analytical instrumentation in the direct-push vehicle on the surface.
The Hydrosparge sampler is lowered through a commercially available direct-push
groundwater sampler that has been advanced into the water table, where it sparges
(purges) VOCs from the groundwater using inert gas; the VOCs are carried to
the DSITMS detector on the
surface for analysis. In this way, groundwater VOCs can be analyzed during advancement
without retrieving the direct-push rods and handling or packaging samples, leading
to increased efficiency and precision and reduced cost over traditional sampling
methods.
TYPICAL USES
As with the MIP,
the Hydrosparge may be useful for characterizing any site with shallow groundwater
VOC contamination (for example, fuel releases, or chlorinated solvent releases
and DNAPL plumes). The system uses a DSITMS detector for sample analysis. As
described previously, DSITMS is a method for the quantitative measurement, continuous
real-time monitoring, and quantitative and qualitative preliminary screening
of VOCs in water, soil, and air.
As with all direct-push
technologies, this system is only applicable to shallow sites with unconsolidated
soils and sediments. The Hydrosparge has been pushed to about 185 feet below
ground surface (bgs) during testing, but 100 feet bgs is considered to be a
realistic average maximum depth.
THEORY OF
OPERATION
The Hydrosparge
system in effect takes a step from the laboratory procedure for VOC analysis
and performs this step in the subsurface. Common methods for VOC analysis, such
as EPA Method
8260B, exploit the volatile nature of VOCs to remove them from their sample
media, either soil or water, into a gaseous phase for introduction into the
analytical instrument. In the case of Method
8260B, VOCs are driven from the sample matrix by passing an inert gas through
it. VOCs are then passed through the analytical detector for quantitative and
qualitative analysis.
The Hydrosparge
system has adapted this approach for in situ analysis. A carrier gas is used
to purge VOCs from the groundwater and transport them to the detector. The Hydrosparge
uses helium as the carrier gas. The system must be closed to ensure that VOCs
are not lost in transport from the surface and that a known quantity of sample
volume is introduced into the detector if quantitative analysis is to be performed.
SYSTEM COMPONENTS
The Hydrosparge
system integrates a customized, 2-inch CPT probe with a small sampling port,
a Teflon transfer line for carrier gas, and an aboveground DSITMS detector in
the truck. The Hydrosparge VOC sensor uses a commercially available HydropunchTM
or PowerpunchTM direct-push groundwater-sampling tool to access the
groundwater. Click here for a diagram of the
system.
MODE OF OPERATION
The HydropunchTM
sampler is pushed to the desired depth and the push rods are retracted, exposing
the screen to the groundwater. The water level is then allowed to come to equilibrium,
which generally takes about 15 to 20 minutes (but this is dependent upon the
hydrogeologic condition). The in situ sparge module (see the link above for
a diagram) is then lowered into the groundwater to operate about 1.5 feet below
the groundwater surface. Using helium gas, the sparge module purges the VOC
analytes in situ from the groundwater to the DSITMS system in the truck, where
VOCs are analyzed in real time.
Click here
to view a short QuickTime video of the Hydrosparge in operation (recommended
for broadband Internet connections).
TARGET ANALYTES
VOCs in groundwater.
PERFORMANCE SPECS
Interferences
The DSITMS does not use a GC for compound separation prior to analysis. Rather, a series
of scans containing ions indicating the presence of VOC analytes is used to
qualitatively identify analytes of interest. Multiple scans of standards containing
the target analytes are integrated, calibrating the instrument to identify these
compounds; however, non-target VOCs that generate an ion with the same mass
and charge ratio as a target analyte may cause a positive interference.
Detection Limits
General detection
limits for DSITMS are typically in the range of 1 ppb or less with little or
no sample preparation required. Sample analysis times are 2 to 3 minutes. Detection
limits for the DSITMS used in conjunction with the Hydrosparge are typically
in the low ppb range, with a linear quantitation range in the low ppm.
Calibration
The Hydrosparge
unit is calibrated by spiking a 250-milliliter flask of distilled water with
known concentrations of VOCs of interest, inserting the in situ-sparge module
into the flask and analyzing the resulting purge gas with the DSITMS. Calibration
of the DSITMS is run initially at start up. A calibration curve is developed
using laboratory prepared standards of known concentrations, bracketing the
expected contaminant concentrations for a particular site. Critical operating
parameters used to generate the calibration curve, including carrier gas flow rate and DSITMS settings, should not be
changed during sample analysis.
Quality Control
Daily calibration,
check standards, and performance evaluation standards are used to ensure data
quality. Confirmation samples may be collected from a groundwater sampler immediately
after removing the in situ sparge module; Method
8260B is recommended for analysis, as this method is quite similar to the
Hydrosparge method.
Precision and
Accuracy
The reliability
of in situ, direct sparging of VOC analytes from groundwater in concert with
the DSITMS has been successfully demonstrated at numerous sites, and the California
Environmental Protection Agency (Cal/EPA) Innovative Environmental Technology
Certification Program has certified the Hydrosparge for use at contaminated
sites. Click here to access the certification notice and study.
According to Cal/EPA:
The technology
was demonstrated to be a qualitative to semi-quantitative field screening
method for TCE, benzene, and carbon tetrachloride and met the criteria of
less than 5% false positives and negatives and had good correlation (R2 =
0.80). For PCE, toluene, and xylenes, the technology was demonstrated to be
a qualitative field screening method and met the criteria of less than 5%
false positives and negatives but had lower correlations (R2 < 0.80). For
DCE, the technology was demonstrated not to meet the criteria of less than
5% false negatives but had good correlation (R2 = 0.80) and could be a qualitative
field screening method for this analyte.
Thermal Desorption Sampler
DESCRIPTION
The
Thermal Desorption VOC Sampler (TDS) is similar in principle and practice to
the MIP and Hydrosparge systems, and is specifically geared toward characterization
of vadose zone soils in situ. The TDS system is a closed system that draws VOCs
directly from the subsurface for analysis by a surface detector. The direct-push
rod is advanced with a special probe that collects a soil plug into a chamber
where it is heated. An integrated pneumatic system transports purged VOCs to
the surface for analysis by DSITMS. The system may be used to collect VOCs onto
analytical traps for later analysis.
The TDS does not
incorporate a lithologic sensor. Sampling intervals are selected based upon
separate pushes with CPT sensors at the desired sample location. As described
previously, standard CPT sensors are deployed in a probe that measures the resistance
against the cone tip and friction against the cone sleeve; system software estimates
the makeup of subsurface soils and sediments by measuring and comparing these
mechanical forces.
TYPICAL USES
As
with the MIP and Hydrosparge systems, the TDS may be useful for characterizing
any site with shallow subsurface VOC contamination from fuel releases, or chlorinated
solvent releases. The TDS is designed for screening on-site soils only, and
uses a DSITMS detector for sample analysis. As described previously,
DSITMS
is a method for the quantitative measurement, continuous real-time monitoring,
and quantitative and qualitative preliminary screening of VOCs.
As with all direct-push
technologies, the TDS system is only applicable to shallow sites with unconsolidated
soils and sediments. CPT probes have been pushed to more than 150 feet bgs during
testing, but 100 feet bgs is considered to be a realistic average maximum depth
for CPT advancement.
THEORY OF OPERATION
Like
the sparge systems, the TDS also takes a step from a laboratory procedure for
VOC analysis and performs this step in the subsurface. Many laboratory and field
analytical methods for VOC analysis, such as EPA
Methods 8260B and 3810,
drive the VOCs from their sample media into a gaseous phase for analysis by
heating the sample. For both methods, an aliquot of sample (either
soil, sediment or water) is removed from the sample vial that has been filled
to eliminate any air pockets. The sample material is placed into a new vial
and heated, driving the VOCs into the headspace of the new vial for removal
and analysis by the detector.
The TDS has adapted
this approach for in situ analysis. As the probe is deployed the sample chamber
is filled, resulting in a sample of known quantity for quantitative analysis.
The sample chamber itself is then heated to increase the mobility of the VOCs
and an inert carrier gas transports them to the surface, where they adsorb onto
an analytical trap. This trap is then heated to drive (desorb) the VOCs into
the DSITMS detector for analysis. The carrier gas system is closed to ensure
that VOCs are not lost in transport from the surface.
SYSTEM COMPONENTS
The TDS consists
of a custom soil probe, carrier gas lines and supply, an analytical trap, and
a DSITMS detector deployed from a direct-push platform. The sample probe incorporates
an internal piston and a heated thermal sample chamber that is connected to
the carrier gas lines (see figure 9 in the following section).
MODE OF OPERATION
The operation of
the TDS is based on the capture of a known volume of soil. The TDS is pushed
to the desired ground depth and an interior rod retracts the penetrometer tip,
which locks into the top of the sample chamber. The probe is then pushed further
into the soil, collecting a 5-gram soil plug in the sample chamber. The soil
plug is heated, releasing the VOC gases from the soil. The vapors are drawn
to the surface by the carrier gas, where they are trapped on an adsorbent media.
The trap is then thermally desorbed into the onboard, field-portable DSITMS,
where VOCs are analyzed in near-real time.
After analysis,
the soil plug is expelled from the sample chamber by reseating the piston into
the drive position, and the sample chamber is heated and purged to remove any
residual contamination before the process is repeated, allowing for screening
of multiple depths during a single push.
Alternatively,
the TDS may be used as a vapor sampler in the vadose zone by applying a vacuum
to the transfer line, drawing soil vapors to the surface where they are trapped,
desorbed, and analyzed by the DSITMS in near-real time.
Figure 9.
 |
For a QuickTime
video of the direct-push TDS system in operation, click here
(recommended for broadband Internet connections).
TARGET ANALYTES
VOCs in vadose
zone soils.
PERFORMANCE SPECS
Interferences
Moisture does not
affect sample results until a sample is saturated, as the sample is dried and
water vapor is purged from the sample prior to analysis. The TDS is designed
to be used in the vadose zone only for this reason.
The DSITMS does
not use a GC for compound separation prior to analysis. Rather, a series of
scans containing ions indicating the presence of VOC analytes is used to qualitatively
identify analytes of interest. Multiple scans of standards containing the target
analytes are integrated to calibrate the instrument to identify these compounds;
however, ions with a similar mass and charge to the target analytes' characteristic
ions could cause interference and artificially increase the reported concentrations.
According to the Cal/EPA certification, the TDS has applicability to field screening
for the presence of known VOCs, and the identification of unknown substances
when ions uniquely characteristic to those substances are present.
The transfer line
is heated to prevent water vapor from condensing in the transfer line, but this
is a potential interference if a large length of line is being used to analyze
a deep sample during very cold conditions.
Detection Limits
The ITMS, using
the conditional EPA Method 8265, is capable of detecting most VOCs qualitatively
and quantitatively in the sub-ppm range. The TDS has been certified as achieving
detection thresholds for trichloroethene (TCE) and total dichloroethene (DCE)
comparable to those of EPA Method 8260A (which has been superceded by EPA
Method 8260B).
Calibration
Calibration of
the DSITMS is run initially at start up. A calibration
curve is developed using laboratory-prepared standards of known concentrations
bracketing the expected contaminant concentrations for a particular site. Critical
operating parameters used to generate the calibration curve, including carrier
gas flow rate and DSITMS settings, should not be changed during sample analysis.
Quality Control
Daily calibration,
check standards, and performance evaluation standards are used to ensure data
quality. Calibration standards should be run daily. Confirmation samples may
be collected by pushing a soil sampler to the same sampling interval screened
by the TDS and analyzed by EPA
Method 8260B.
Precision and
Accuracy
The reliability
of in situ, thermal desorption of VOCs from the soil, in concert with the DSITMS
has been successfully demonstrated at various sites. A number of field demonstrations
have provided direct comparisons between the TDS and DSITMS combination, and
the standard EPA-mandated procedures. Field studies for total DCE and TCE demonstrated
that the TDS technology achieved less than 5 percent false negative results
and less than 5 percent false positive results when compared to verification
core samples analyzed by EPA
Method 8260B. For the two analytes having the most performance data there
appears to be a relatively good correlation between the TDS sample results and
verification sample result (r2 =0.83 for DCE; r2 =0.97 for TCE). Cal/EPA has
certified the TDS as a near real-time, qualitative to semi-quantitative, in
situ subsurface field screening method for VOCs in the vadose or capillary zone.
Click here to access the certification notice and study.
Inorganic
Detectors
DESCRIPTION
While much attention has been paid to analytical systems for organic analysis,
two systems have been developed for deployment on direct-push platforms for
screening inorganic contaminants in the subsurface. The first has adapted XRF
technology to subsurface characterization. XRF is a well-established, non-destructive
laboratory and hand-held field screening method for determining elemental concentrations
at ppm levels. The second system has been built around a more recent technology,
laser induced breakdown spectroscopy (LIBS). Both systems are used in the detection,
identification, and delineation of heavy metal contaminants in the subsurface.
The LIBS system operates in the unsaturated and capillary zones, while the XRF
system may be applied in both the unsaturated and saturated zones.
TYPICAL USES
Direct-push inorganic
systems are applicable to a variety of sites with shallow heavy metal contamination
in unconsolidated sediments, due to their multiple analyte detection capability.
Direct-push deployment of metals sensors allows for high-resolution delineation
of subsurface metals contamination, with no waste generation.
THEORY OF
OPERATION
XRF
In XRF analysis, a process known as the photoelectric effect is used in analyzing samples. Fluorescent x-rays are produced by
bombarding the subsurface soils with an x-ray source that has an excitation
energy similar to, but greater than, the binding energy of the inner-shell electrons
of the elements of interest present in the soils. Some of the source x-rays
will be scattered, but a portion will be absorbed by these target elements.
Because of their higher energy level, they will cause ejection of the inner-shell
electrons, and the electron vacancies that result will be filled by electrons
cascading in from outer electron shells. However, since electrons in outer shells
have higher energy states than the inner-shell electrons they are replacing,
the outer shell electrons must give off energy as they cascade down. The energy
is given off in the form of x-rays, and the phenomenon is referred to as x-ray
fluorescence (click to view a schematic diagram
of the x-ray fluorescence process). Because every element has a different electron
shell configuration, each element emits a unique x-ray at a set energy level
or wavelength that is characteristic of that element. The elements present in
a sample can be identified by observing the energy level of the characteristic
x-rays, while the intensity of the x-rays is proportional to the concentration
and can be used to perform quantitative analysis. In other words, qualitative
analysis is performed by observing the energy of the characteristic x-rays.
A quantitative analysis is performed by measuring the intensity of the x-ray.
LIBS
The LIBS sensors use high-power pulsed lasers to generate plasma in the soil.
The output of the laser beam is focused on the surface of the soil outside the
probe. This causes a breakdown of the soil and the formation of a high temperature
plasma spark. For a brief time, this plasma spark emits light.
The wavelengths of the light, or the constituent
colors, are indicative of the elements present in the soil; specific wavelengths
correspond to specific metals. The brightness of the light at a metal's wavelength
indicates how much of that metal is present. A spectrometer breaks
this light into its constituent colors, much like the action of a prism. This
information is analyzed onboard the direct-push rig to obtain qualitative and
quantitative data from the characteristic signatures from each of the metals
detected. An example of the readout is presented below in Figure 10. The analysis
is similar to that performed by an inductively coupled plasma analyzer. The
advantage of LIBS is that little or no sample preparation is required to obtain
useful results and the technique is readily portable to the field.
Figure 10.
 |
For a more detailed
discussion of the theory behind XRF and other applications of the technology,
visit the XRF section.
SYSTEM COMPONENTS
XRF
The
XRF system shown in Figure 11 consists of a probe containing an x-ray tube configured
to focus its beam through a boron carbide window, a detector and preamplifier,
a collimator, a boron carbide window, and standard
CPT sensors for soil lithology and moisture. A high-voltage cable delivers power
down the inside of the direct-push tool string to the x-ray tube, and a detector
cable returns the signal to the multichannel analyzer and computer in the truck
on the surface.
Figure 11.
|
LIBS
| Figure 12. |
 |
The fiber-optic LIBS
system (Figure 12) is configured with a laser in the direct-push rig, a fiber-optics
transmission system extending down the inside of the tool string, and a custom
probe combining the lens assembly for the LIBS laser with standard CPT sensors
for soil lithology and moisture. The downhole LIBS configuration houses the laser
and fiber-optic components within the probe, along with standard CPT sensors.
Standard CPT sensors
measure the resistance against the cone tip and friction against the cone sleeve;
system software estimates the makeup of subsurface soils and sediments by measuring
and comparing these mechanical forces.
Click for a schematic of the complete fiber-optic LIBS system or the downhole LIBS.
MODE OF OPERATION
XRF
The XRF sensor is advanced to a selected sampling depth at which point an x-ray
source in the probe tip bombards the surrounding soil. The source may be either
an x-ray tube or a radioisotope. X-ray tubes are more powerful and achieve lower
detection limits, but a radioisotope source generates a more consistent output
and may deliver more precise and accurate results. The operator must take care
and follow all applicable regulations when shipping and handling a radioisotope
source. Radioactive materials handling licenses may be required by the state
or locality in which it is used.
Metal atoms present in the soil are excited and emit fluorescent x-rays with
an energy that is characteristic for the specific elements. These emitted x-rays
are detected at the probe tip and provide an individual peak for each type of
metal present in the soil. These signatures are identified and quantified in
real time onboard the direct-push rig.
For a QuickTime video of the direct-push XRF system in operation, click
here (recommended for broadband Internet connections).
LIBS
The LIBS probe is advanced into the subsurface with standard direct-push tooling.
The output of the laser beam is focused on the surface of the soil outside the
probe and emits periodic pulses during advancement, causing a breakdown of the
soil and the formation of a high temperature plasma spark. The wavelengths of
light (indicative of the elements present in the soil) are picked up by a detector
in the probe that relays the electronic signal to the spectrometer on the surface
in the truck, which breaks this light into its constituent colors, much like
the action of a prism. This information is analyzed onboard the direct-push
rig to obtain qualitative and quantitative data from the characteristic signatures
from each of the metals detected. The resulting LIBS data are used to generate
detailed graphics depicting the metals concentration as a function of depth.
The LIBS system also can be deployed in a stand-alone system, without the CPT
but with a backpack or cart-mounted system, to analyze surficial soil samples
or grab samples. The unit is designed to be used above ground, and proper care
must be taken to prevent accidental discharges between probes and during handling.
Operators should wear appropriate skin and eye protection during use.
TARGET ANALYTES
Heavy metals
PERFORMANCE SPECS
Interferences
Matrix effects
can cause a great deal of variation in inorganic sample analyses. Physical matrix
effects result from variations in the physical character of the sample soils,
such as particle size, uniformity, homogeneity, and condition of the surface.
Metals analyses by both XRF and LIBS are especially susceptible to heterogeneity
of contaminant distribution in the soil, which is a common problem with inorganic
contaminants. The contaminants are not removed from the matrix as with formal
laboratory procedures. The option of homogenizing the sample is not available
for in situ analysis.
There
are several other interferences that can affect the ability of these sensors
to detect and quantify elements in a sample. Moisture content above 20 percent
may cause problems for XRF analysis (since moisture alters the soil matrix for
which the XRF has been calibrated), but these problems are manageable and XRF
may be used in the saturated zone. By contrast, LIBS is very sensitive to moisture
and does not perform well in saturated soils and sediments. Because typical
x-ray penetration depths range from 0.1 to 1 millimeter (mm), the window of
the XRF probe should be in direct contact with the sample. Chemical matrix effects
also can hamper XRF analysis as x-ray
absorption and enhancement phenomena. Finally, the resolution of the detector
may cause problems in analyzing some elements. If the energy difference between
the characteristic x-rays of two elements (as measured in eV)
is less than the resolution of the detector in eVs, the detector will not be
able to resolve the peaks. In other words, if two peaks are 240 eVs apart, but
the resolution of the detector is 270 eV, the detector will have difficulty
in differentiating those peaks.
Detection Limits
XRF can detect heavy metals at levels below 100 ppm, up to the full depth allowable
by the direct-push rig. A field test of the LIBS system at a national laboratory
successfully measured a range of chromium concentrations from 30 ppm (background)
to 1,200 ppm. In general, LIBS is a more sensitive technology than XRF and can
achieve lower detection limits, often in the single ppm range. Laboratory tests
of the LIBS system were used to perform quantitative analysis of lead in soils.
From this data, it was determined that the minimum LIBS detection limit for
lead in soils was 10 to 40 ppm.
Calibration
The
XRF is calibrated by analyzing standard reference materials (SRM) from the National
Institute of Standards and Technology containing varying concentrations
of the analytes of interest. Additional calibration standards may be created
by spiking clean soils with known concentrations of one or more of the target
analytes at concentrations representative of what is expected at the site. The
calibration samples are placed in a sample chamber attached to the sample probe,
which protects the operator from x-ray exposure during analysis. The calibration
samples are measured with three to five replicates each for a 100-second acquisition
time. A calibration curve
is then created by performing a linear
least-squares fit to the SRM and prepared samples if used.
The LIBS is typically
calibrated in the laboratory prior to going into the field. A site-specific
suite of calibration standards is prepared by spiking clean soils from the site
with known concentrations of standards. Using site soils allows the operator
to tailor the calibration to the soil conditions found on site. A "zero"
sample of clean soil is analyzed along with prepared samples containing a range
of contaminant concentrations expected to be encountered at the site. A "nominal"
calibration consisting of a software generated generic calibration may be performed
if relative response data is acceptable and an accurate lower detection limit
is not necessary.
Quality Control
XRF and LIBS are field screening methods and do not eliminate the need for traditional
laboratory analyses. Soil samples must be collected from an adjacent location
at a percentage (typically 5 to 10 percent) of field screening locations and
submitted to an off-site laboratory for analysis. These results are used to
recalibrate the sensor in the case of XRF, and to confirm that the sampling
results are accurate in both cases. Field data may be mathematically corrected
after the fact based upon the confirmatory sample results if a clear trend is
apparent for a given analyte or analytes. Calibration check samples may be analyzed
by retrieving the XRF probe from the soil periodically and re-analyzing a calibration
standard. The purpose of the calibration check is to assess whether the instrumentation's
response has remained constant during operation.
Precision and Accuracy
During the previously-described
field test of the LIBS system, measurements of chromium concentrations from
30 ppm (background) to 1,200 ppm correlated highly with data collected from
other soil borings taken in the test location.
A report containing more detailed information about
the XRF and LIBS may be found by following this link.
Explosives Sensor
DESCRIPTION
The Explosive Sensor (ES) probe detects total concentrations
of explosives contamination in the subsurface soil. The ES uses electrochemical
sensors that detect the presence of certain chemical compounds characteristic
of explosive compounds. The probe also incorporates geotechnical sensors (tip
resistance and sleeve friction sensors) for determining soil lithology. By combining
the ES and geotechnical sensors, the probe collects soil classification and contaminant
concentration versus depth information during the penetrometer push. The ES
is currently in the precommercial stage, and is available for adoption by eligible
agencies of the Federal government or for commercialization.
The probe incorporates an external pyrolyzer system used to transform explosive
compounds into electroactive vapors and a pneumatic system to transport these
vapors from the soil to the electrochemical sensors inside the probe. Various
other materials present in the soil may contain some of the same compounds characteristic
of explosives. Although unable to differentiate between specific explosives
compounds, the ES is equipped with electrochemical sensors to differentiate
between compounds containing organic nitrogen and inorganic nitrogen, and to
distinguish explosives compounds from other compounds, such as common fertilizers.
A diagram of the system is provided below in Figure 13.
Figure 13.
 |
TYPICAL USES
The ES is used
for assessment of sites contaminated with energetic materials such as the explosives
trinitrotoluene (TNT), cyclotrimethylenetrinitramine (RDX), Cyclotetramethylenetetranitramine
(HMX), their manufacturing intermediates, and subsequent breakdown products.
Characterization of these sites is often very time- and cost-intensive because
contaminant distribution is generally widespread over large areas. Ammunition
plants typically cover many thousands of acres. The traditional methods to measure
TNT and other explosives in the environment include collection, transportation
of samples, and extraction, coupled with sophisticated laboratory analysis (Yinon
and Zitrin 1981). Direct-push analytical instrumentation offers cost relief
and time savings for such large-scale sites.
As with all direct-push
technologies, the ES is only applicable to shallow sites with unconsolidated
soils and sediments. CPT probes have been pushed to over 150 feet bgs during
testing, but 100 feet bgs is considered to be a realistic average maximum depth
for CPT advancement. Furthermore, the ES may only be used to characterize vadose
zone contamination.
THEORY OF
OPERATION
Explosive compounds such as TNT can be thermally decomposed into vapor.
In order to decompose explosives in situ, the soil is heated to release the
nitrogen-containing gases that are characteristic of explosives. The analyte
vapor must be electrochemically active and must come into physical contact with
the active electrode of the sensor. The nitric oxide (NO) gas sensor cannot
directly measure a response from solid TNT because of its low vapor pressure.
For this reason, explosive compounds must be vaporized from the soil matrix
for analysis.
Vaporization is
achieved using a heater located on the outside of the probe, in close proximity
to the sensor, to thermally degrade TNT and other explosives that are generally
found in crystalline form in the soil. During operation, the wire is heated
briefly to about 900°C. The surrounding soil (from 0.15 to 0.25 centimeter
[cm] from the filament) is warmed from 125°C to 150°C. The explosives
in the soil sublime, and the heat dissociates the NO2 groups that
are weakly bound to the main portion of energetic molecules. The resulting vapor
contains carbon monoxide (CO), carbon dioxide (CO2), water (H2O)
and nitrous oxide (NO2). The NO2 molecules are further
dissociated into NO and oxygen. The NO fragment, nitric oxide, is the dominant
gaseous product.
The electrochemical
sensor measures a change in current as result of contact by the NO2 ion with
the sensor electrode. The voltage response is a direct function of the total
concentration of explosive compounds.
SYSTEM COMPONENTS
The explosives
probe contains, in addition to standard CPT geotechnical sensors, an electrochemical
sensor that is responsive to NO2, an internal pneumatic system, associated power
supply and signal conditioning electronics inside the probe, and output monitors
in the truck. The probe is designed for air to be continuously pumped through
the output ports of the probe and into the soil matrix adjacent to the probe,
collected through the vapor inlet ports, directed over the sensor, and drawn
to the surface where it is vented to the outside.
A CO sensor is
included in the probe design. Multiple sensors permit discrimination of organic
nitroaromatic compounds from inorganic nitrogen compounds that do not give off
CO compounds when dissociated by heat. The heater consists of a 20-cm length
of 0.010-inch-diameter platinum wire wound in two loops around a ceramic inset
in the probe. It is located between the output and inlet ports so that the heater
wire is offset from the soil surface by 0.15 inch. A commercial zirconium oxide
ceramic paint is used as an electrical insulator between the platinum wire and
the probe's surface.
MODE OF OPERATION
The probe is pushed
to the full depth of interest at a constant rate of 2 cm per second to collect
soil stratigraphy data using the CPT sensors. After pushing to the maximum depth,
the probe is retracted about a foot, releasing a sacrificial sleeve that protects
the pyrolyzer unit during the downward push, exposing the wire heater, the module's
vapor delivery, and the sampling ports. Sensor readings are then made at discrete
depths during retraction. Each reading consists of a short stabilization period
of 0.5 to 1 minute, a 30-second activation of the heating elements to thermally
degrade explosives in the surrounding soil, followed by another stabilization
period of about 30 seconds while the sensor recovers. The probe is then retracted
to the next sampling point. The probe may be pushed to the full depth allowed
by the umbilical cable length. The minimum depth at which the sensor is able
to operate is 4 inches bgs.
TARGET ANALYTES
Explosives (TNT,
RDX, HMX, their manufacturing intermediates, and subsequent breakdown products)
in the vadose zone.
PERFORMANCE
SPECS
Interferences
An electrochemical
CO sensor is used with the ES to differentiate between compounds containing
organic nitrogen and inorganic nitrogen, and to distinguish explosives compounds
from other compounds, such as common fertilizers.
Matrix interferences
are a concern because explosives are relatively insoluble crystalline compounds
that tend to be dispersed very heterogeneously. Solid TNT is generally not uniformly
distributed but instead exists in soils as localized particles. Under these
circumstances, samples collected from nearly identical site locations and depths
can have significantly different contamination levels.
Detection Limits
Laboratory results
indicated that the probe is very sensitive to low concentrations of TNT in dry
soil. The calculated lower detection limit based on the signal equivalent to
three times the noise is 0.5 ppm.
Calibration
The standard method
of operating the ES in the field consists of calibrating the stratigraphy, electrochemical
sensors, and air flow prior to pushing the probe. A known concentration of NO
in nitrogen gas is used to calibrate the explosives electrochemical sensor.
Quality Control
After initial sensor
measurements indicate explosives contamination in a particular area, a direct-push
soil sampler is used to obtain 18-inch cores at depths suspected of having explosives
contaminants. The verification soil samples may be analyzed in the field using
an immunoassay field
test procedure or sent to an off-site laboratory for confirmatory analysis by
EPA Method 8330.
Precision and Accuracy
The response of
the probe is linear over a broad range of concentrations. Although not exact,
the explosives sensor probe readings correspond well with the laboratory results
obtained by EPA
Method 8330 and the immunoassay field method. Some of the discrepancies
between the three methods may be attributed to the heterogeneity of explosives
in environmental samples. The ES is designed to be a rapid site screening tool
to distinguish areas of contamination from areas without contamination, and
functions as a semi-quantitative tool.
A report containing
more detailed information about the ES may be found by following this link.
Advantages
Direct-push analytical
systems share a number of advantages over traditional investigative methods,
including drilling and collection of samples for off-site analysis (regardless
of the sampling platform). These advantages are inherent to in situ analytical
systems and direct-push platforms. In addition, the individual analytical systems
described above have their own particular strengths. The following advantages
apply to some or all of the systems described above.
- Generation
of (near) real-time data, which often allows completion of site characterization
in one mobilization
- Investigation
flexibility due to real-time data generation
- Greater throughput than traditional sampling methods, which allows for
a greater amount of data to be generated and overall site uncertainty to
be minimized
- The ability to access hard-to-reach or sensitive areas, or areas with limited
overhead clearance (for example, power lines, inside buildings)
- Generation of little to no investigation-derived waste
- Vertical sampling capabilities to about 60 feet bgs for rotary hammer
rigs and 100 feet bgs for CPT rigs
- The ability to characterize subsurface contamination at multiple intervals
(or continuously) during one push (Hg fluorescence, MIP, TDS, LIBS, XRF)
- Simultaneous generation of lithologic logs (MIP)
- Sampling capabilities in both unsaturated and saturated zones (MIP, XRF)
Limitations
Direct-push analytical
systems have limitations that are important to keep in mind when considering their
application for site characterization, including:
- Detection limits
can be higher and precision and accuracy lower than with traditional analytical
methods
- Some systems
may be semi-quantitative (Hydrosparge and MIP in unsaturated soils) or semi-qualitative
(MIP/PID and MID/FID configurations and Hg fluorescence) depending on the
target contaminants and applications
- Because experienced
operators are generally required, it may not be possible to lease the systems;
many may be offered as contract services only
- Practical vertical
sampling limits are about 100 feet bgs for rotary hammer rigs and 150 feet
bgs for CPT rigs
Cost Data
Studies
indicate that direct-push analytical systems may provide significant savings
over conventional site assessment and characterization methods. Studies performed
by the U.S. Department of Energy for CPT-mounted systems indicate that 25- to
35-percent savings may be achieved. Cost information varies greatly among the
different technologies as well as for projects of different scope. The following
contractors and vendors may be contacted for more detailed cost assessments.
Applied Research Associates
Geoprobe Systems
Gregg Drilling
Additional Resources
Using The Triad
Approach To Improve The Cost-Effectiveness Of Hazardous Waste Site Cleanups
Tri-Service
Site Characterization and Analysis Penetrometer System (SCAPS) Accelerated Sensor
Development Project - Final Report
Verification/Evaluation
Reports
Verification of the performance of site characterization and field analytical
technologies is conducted through a variety of programs. Evaluation and verification
reports for several of the technologies described in this encyclopedia entry
are provided below.
The reliability of in situ, direct sparging of VOC analytes from groundwater
in concert with the DSITMS has been successfully demonstrated at numerous sites,
and Cal/EPA has certified the Hydrosparge for use at contaminated sites. Click here to
access the certification notice and study.
Cal/EPA has certified the TDS as a near real-time, qualitative to semi-quantitative,
in situ subsurface field screening method for VOCs in the vadose or capillary
zone. Click here to access the certification notice and study.
Click
here to download QuickTime
video player.
Disclaimer
http://www.cluin.org/char/technologies/dpanalytical.cfm
Page Last Modified: May 18, 2007

|