This page uses javascript to control a rollover affect of linked images. No functionality is lost with javascript turned off.
CLU-IN HomeSkip common site navigation and headers
What's Hot? What's New? Remediation Characterization and MonitoringTraining Initiatives and Partnerships Publications and Studio Databases Software and Tools TechDirect and Newsletters Vendor and Developer Support About CLU-IN

Technology Innovation Program
CommentsSite MapEPA HomeCLU-IN Home
  Characterization and Monitoring  
  - Technology Tools
> Technology Descriptions and Selection Tools
 
 
Select a Technology | Field Analytic Technologies Index | Ask an Expert | Problem Solver  

X-ray Fluorescence

XRF Instrument

Description

Field-portable, handheld device for simultaneously measuring a number of metals in various media.

Typical Uses

Energy dispersive X-ray fluorescence (EDXRF) is a method of detecting metals and other elements, such as arsenic and selenium, in soil and sediment. Some of the primary elements of environmental concern that EDXRF can identify are arsenic, barium, cadmium, chromium, copper, lead, mercury, selenium, silver, and zinc. Field-portable X-ray fluorescence (FPXRF) units that run on battery power and use a radioactive source were developed for use in analysis for lead-based paint and now are accepted as a stand-alone technique for lead analysis. In response to the growing need for field analysis of metals at hazardous waste sites, many of these FPXRF units have been adapted for use in the environmental field. The field-rugged units use analytical techniques that have been developed for analysis of numerous environmental contaminants in soils. They provide data in the field that can be used to identify and characterize contaminated sites and guide remedial work, among other applications.

More recently, FPXRF analyzers have been used to detect metals in water. The water samples must be filtered and concentrated with an ion exchange membrane to achieve detection limits in the low parts per billion (ppb) range lower than applicable maximum contaminant levels (MCL). Many manufacturers of FPXRF units currently are conducting research to refine the procedures for preparation of water samples to make FPXRF analysis a practical field analytical technique for metals in water.

Theory of Operation

FPXRF instruments were developed as an effective and nondestructive tool for measuring lead in paint and in house dust. Most homes constructed or painted before 1976 contain lead-based paint, which is one of the most common sources of lead ingested by children. In response, the U.S. Department of Housing and Urban Development (HUD) set guidelines for the inspection and abatement of contamination in public housing developments at which lead paint had been used. HUD considers any paint surface with a lead content greater than 1 milligram per square centimeter (mg/cm2) to be a lead-based paint surface. FPXRF units were designed to detect lead in paint at levels at or lower than that level. Air filters are used to measure concentrations of metals in household dust. When the volume of air that has passed through the air filter is measured, a conversion can be made to determine the concentrations of metals suspended as particulates in the air. Although the technique was developed for homes in which contamination with lead is suspected, it also has been applied in monitoring air emissions from industrial processes or from remediation processes conducted at a hazardous waste site.

One of the advantages of EDXRF analysis is that it can be used not only to detect lead, but also to detect and measure many elements simultaneously. Generally, EDXRF units can detect and quantify elements from atomic number 16 (sulfur) through 92 (uranium). There are two types of EDXRF units: bench-top units that use an X-ray tube source and FPXRF analyzers that use a radioisotope as a source of X-rays. Instruments that use X-ray tubes as sources commonly are not used in the field because of the larger power requirements for the X-ray tube and the added weight of the instrument. Use of a radioactive source eliminates the need for a fixed power source for an X-ray tube, making the FPXRF unit truly portable.

In FPXRF analysis, a process known as the photoelectric effect is used in analyzing samples. Fluorescent X-rays are produced by exposing a sample to an X-ray source that has an excitation energy similar to, but greater than, the binding energy of the inner-shell electrons of the elements in the sample. Some of the source X-rays will be scattered, but a portion will be absorbed by the elements in the sample. Because of their higher energy level, they will cause ejection of the inner-shell electrons. The electron vacancies that result will be filled by electrons cascading in from outer electron shells. However, since electrons in outer shells have higher energy states than the inner-shell electrons they are replacing, the outer shell electrons must give off energy as they cascade down. The energy is given off in the form of X-rays, and the phenomenon is referred to as X-ray fluorescence (click to view a schematic diagram of the X-ray fluorescence process). Because every element has a different electron shell configuration, each element emits a unique X-ray at a set energy level or wavelength that is characteristic of that element. The elements present in a sample can be identified by observing the energy level of the characteristic X-rays, while the intensity of the X-rays is proportional to the concentration and can be used to perform quantitative analysis. In other words, qualitative analysis is performed by observing the energy of the characteristic X-rays. A quantitative analysis is performed by measuring the intensity of the X-ray.

The emissions of characteristic X-rays from three electron shells commonly are involved in FPXRF analysis: the K, L, and M shells. A typical emission pattern, or emission spectrum, for a given element has several peaks generated from the emission of X-rays from those shells.

System Components

A FPXRF system has two basic components: the radioisotope source and the detector. The source irradiates the sample to produce characteristic X-rays, as described above. The detector measures both the energy of the characteristic X-rays that are emitted and their intensity to identify and quantify the elements present in the sample. The following sections describe each of the components in greater detail.

Radioisotope Sources

An X-ray source will excite characteristic X-rays from an element only if the source energy is greater than the binding energy, or absorption edge energy, of the electrons in a given electron shell. A given individual source can analyze only certain elements. Analysis is more sensitive for an element with an absorption edge energy similar to, but less than, the excitation energy of the source. For example, when using a cadmium-109 (C-109) source, FPXRF would exhibit more sensitivity to zirconium, which has a K shell energy of 15.7 kiloelectron volts (keV), than for chromium, which has a K shell energy of 5.41 keV.

The radioisotope sources that are becoming standard in FPXRF units are Fe-55, Cd-109, and Am-241. Elements that those sources commonly analyze include:

  • Fe-55: sulfur (S), potassium (K), calcium (Ca), titanium (Ti), and chromium (Cr)
  • Cd-109: vanadium (V), Cr, manganese (Mn), iron (Fe), cobalt (Co), nickel (Ni), copper (Cu), zinc (Zn), arsenic (As), selenium (Se), strontium (Sr), zirconium (Zr), molybdenum (Mo), mercury (Hg), lead (Pb), rubidium (Rb), and uranium (U)
  • Am-241: cadmium (Cd), tin (Sn), antimony (Sb), barium (Ba), and silver (Ag)

Because individual sources by nature reliably analyze only a limited number of sources, FPXRF instruments that use more than one source have been developed, allowing them to analyze a greater number and range of elements. Typical arrangements of such multisource instruments include Cd-109 and Am- 241 or Fe-55, Cd-109, and Am-241.

X-ray Tube Sources

Miniature x-ray tube sources are now being employed by a number of vendors. The advantage of the x-ray tube sources is that it does not require licensing or special shipping, as do XRF units employing radioactive sources. These units usually have a low-power hot-filament cathode x-ray tube. The transmission anode operates at a high enough energy range (~35 keV) in order to simultaneously excite a large range of elements (k through u). Interferences and sensitivity problems associated with high energy sources are corrected using sophisticated software built into the XRF unit.

Detectors

Two basic types of detectors are used in FPXRF units: gas-filled and solid-state. Each detector has its advantages and limitations and is better suited to some applications than to others.

Common solid-state detectors include Si(Li), HgI2, and silicon pin diode. Among those detectors, the Si(Li) is capable of the highest resolution but is quite temperature-sensitive and will register signal "noise" if not cooled sufficiently. The Si(Li) has a resolution of 170 electron volts (eV) if cooled to at least –90°C, either with liquid nitrogen or by thermoelectric cooling that uses the Peltier effect. The HgI2 detector can operate at a moderately subambient temperature and is cooled by use of the Peltier effect. It has a resolution of 270 to 300 eV. The silicon pin diode detector operates near ambient temperatures and is cooled only slightly by use of the Peltier effect. It has a resolution of 250 eV.

Some elements produce peaks that are near each other in the spectrum, while very high concentrations of one element may produce a peak that overwhelms the peaks of other elements that are present at lower concentrations. The higher the resolution, the better the detector is able to separate characteristic peaks. The XRF operator must be careful to select an FPXRF unit that has sufficient resolution to satisfy the data quality needs of the project. The following link provides an illustration of this concept by providing the resolution differences among some common XRF detectors. Resolution is discussed in greater detail in a later section.

Mode of Operation

The radioisotope source or sources are housed in a metal turret, with additional lead shielding inside the probe. To perform an analysis, a sample is positioned in front of the plastic film measurement window of the probe and measurement of the sample is initiated, usually by depressing a trigger or start button. Doing so exposes the sample to the source radiation. For units that use multiple sources, after the sample has been exposed to one source, the turret is rotated to expose it to the next source. The length of time the sample actually is exposed to each source is referred to as the count time. The sample is exposed to the radioactive source for a number of seconds. Fluorescent and backscattered X-rays from the sample reenter the analyzer through the window and are counted by the instrument's detector. X-rays emitted by the sample at each energy level are called "counts". The detector records the counts, measures the energy of each X-ray and builds a spectrum of analyte peaks on a multichannel analyzer (MCA). The unit’s software integrates the peaks to produce a readout of concentrations of analytes, and, usually, the standard deviation for each analyte. Numerous sample results and spectra can be stored for later viewing, downloading into a computer, or printing. Some units also allow the operator to recall previous results and even to view their spectra. At the completion of the exposure time, the instrument software statistically computes a concentration from the readings collected from each energy level along the spectrum. Count times are not to be confused with the total analytical time, which includes all of the analytical functions, such as rotation of the source into position, and processing of the results by the instrument software, in addition to the count time of each source.

Count times from 30 seconds per source to as long as 200 seconds per source can be employed, depending on the data quality needs of the project. As count times increase, the detector collects a larger number of X-rays from the sample, including more X-rays from elements that are present at comparatively lower concentrations. For that reason, the longer the count time, the lower the detection limits; typically, quadrupling the count time will cut the detection limit in half. For example, if a 50-second count time yields a detection limit of 100 parts per million (ppm) for a given element, increasing the count time to 200 seconds will lower the detection limit to approximately 50 PPM Using the instrument's software, the operator can select the appropriate count times.

An FPXRF detector can be operated in the in situ or the intrusive mode. Count times of 30 to 60 seconds per source are common for in situ analysis, while count times for intrusive analysis may be as long as 200 seconds per source. The particular requirements of the job, such as the required detection limits or data sample precision, and the purpose of sampling--for field screening or for definitive analysis--will determine which mode is appropriate and what count times are needed. Descriptions of each mode follow.

XRF Instrument In situ analysis refers to the rapid screening of soils in place. For in situ operation, the window of the probe is placed in direct contact with the surface to be analyzed, and a trigger is pulled, much as one would fire a gun. Because analyses in this mode typically are completed very quickly (in less than one minute) and sample heterogeneity of the samples sometimes is a concern, it is recommended that three to four measurements be taken in a small area and the values be averaged to determine the concentrations of metals. Intrusive Operation Intrusive analysis used to ensure greater precision when lower detection limits are needed. Those goals are achieved through more extensive sample preparation and longer analysis times to reduce heterogeneity among samples and increase the sensitivity of the instrument, respectively. For intrusive operation, a sample is collected, prepared (usually by homogenizing, drying, grinding, and sieving), and placed in a 31- or 40-millimeter (mm) polyethylene sample cup that has a transparent Mylar window. The sample cup is placed over the probe window (some units provide a safety cover for intrusive analysis) and analyzed. Some FPXRF instruments can analyze samples in either mode, while others have only one mode of operation.Standard Operating Procedures (SOPs)

While a clear distinction is made here between in situ analysis and fully intrusive analysis, sample analysis is in reality a continuum. Thorough homogenization will improve the precision and accuracy of the analysis dramatically; an "in situ prepared" sample can be collected, homogenized, and analyzed right next to the sample location (possibly right through a plastic bag used for homogenization). Drying the sample also may improve the results significantly, and, depending on the project’s data quality objectives, homogenization and drying may be all the preparation required for an intrusive analysis. Preparation of samples is discussed in greater detail in a later section.

Target Analytes

The target analytes are metals and other nonmetallic elements, such as arsenic and selenium.

Performance Specs

Performance specifications include information about interferences, detection limits, calibration, sample preparation, quality control, and precision and accuracy.

Interferences

There are a number of factors known as interferences that can affect the detection and quantification of elements in a sample. Some interferences can be inherent in the method of analysis, while others are the result of the instrument's setup, such as calibration methods. Other interferences may arise from outside sources, such as the sample matrix (for example, soils and sediment). Some factors can be prevented or minimized through careful preparation and sample design; others are natural effects that must be taken into consideration. To produce useful data, it is important that the analyst understand the interferences. Their effects and the procedures used to evaluate them are described below.

Matrix Effects
Matrix effects can cause a great deal of variation in sample analyses. Physical matrix effects result from variations in the physical character of the sample soils, such as particle size, uniformity, homogeneity, and condition of the surface. The FPXRF demonstration conducted under the Superfund Innovative Technology Evaluation (SITE) program provided convincing evidence that the heterogeneity of the sample generally has the greatest effect on comparability with confirmatory samples. Every effort should be made to homogenize soil samples thoroughly before analysis. One way to reduce particle size effects is to grind and sieve all soil samples to a uniform particle size.

Moisture Effects
Moisture content above 20 percent may cause problems, since moisture alters the soil matrix for which the FPXRF has been calibrated. This problem can be minimized by drying, preferably in a convection or toaster oven. Drying by microwave can increase variability between the FPXRF data and confirmatory data and can cause arcing if fragments of metal are present in the sample.

Sampling Effects
In environmental samples, typical X-ray penetration depths range from 0.1 millimeter (mm) to 1 mm. Inconsistent positioning of samples in front of the probe window is a potential source of error because the X-ray signal decreases as the distance from the radioactive source increases. Maintaining a consistent distance between the window and the sample minimizes that problem. For best results, the window of the probe should be in direct contact with the sample.

Chemistry Effects
Chemical matrix effects also can occur as X-ray absorption and enhancement phenomena. For example, iron tends to absorb copper X-rays, while chromium actually will be enhanced in the presence of iron. The effects can be corrected mathematically through the FPXRF instrument’s software.

Detector Resolution Effects
The resolution of the detector may cause problems in analyzing some elements. If the energy difference between the characteristic X-rays of two elements (as measured in eV) is less than the resolution of the detector in eVs, the detector will not be able to resolve the peaks. In other words, if two peaks are 240 eVs apart, but the resolution of the detector is 270 eV, the detector will have difficulty in differentiating those peaks. A common example is the overlap of the arsenic K peak with the lead L peak. With the use of mathematical corrections that subtract the lead interference, lead can be measured from the lead L peak and arsenic still can be measured from the arsenic K peak. However, concentrations of arsenic cannot be calculated efficiently for samples that have lead to arsenic ratios of 10 to 1 or more, because the lead peak will overwhelm the arsenic peak completely.

Detection Limits

An FPXRF operator must consider two types of detection limits: instrument detection limits (DL) and method detection limits (MDL). A DL is the absolute threshold concentration of a given element that a particular instrument can resolve, as determined by the standard deviation (SD) of an individual analytical result. DLs of 10 to 100 PPM are typical for soil samples, although DLs may be higher for elements like chromium and cadmium that have characteristic X-ray peaks far removed from the energy level of the sources typically used.

MDLs depend on the analytical method (such as preparation and analysis times) and may be higher than DLs. The results of replicate measurements of a low-concentration sample can be used to generate an average site-specific MDL. The MDL is defined as three times the SD of the results for a replicate analysis of a low-concentration sample. With the exception of chromium which has a MDL as high as 900 milligram per kilogram (mg/kg) depending the instrument being used, the MDLs for most analytes are in the range of 40 to 200 mg/kg.

Click to view a comparison of method detection limits for six commercial FPXRF instruments.

Calibration

FPXRF units are calibrated by any of several methods. The methods will vary according to the make of the unit and the use to which the data are to be put, such as for screening or for definitive analysis. Basically, there are two types of calibration, although there is some overlap between two.

Fundamental Parameters Calibration
The fundamental parameters (FP) calibration is a “standardless” calibration. Rather than calibrating a unit's calibration curve by measuring its response to standards that contain analytes of known concentrations, FP calibration relies on the known physics of the spectrometer's response to pure elements to set the calibration. Built-in mathematical algorithms are used to adjust the calibration for analysis of soil samples and to compensate for the effects of the soil matrix. The FP calibration is performed by the manufacturer, but the analyst can adjust the calibration curves (slope and y-intercept) on the bases of results of analyses of check samples, such as standard reference materials (SRM), which are analyzed in the field.

Empirical Calibration
In performing an empirical calibration, a number of actual samples, such as site-specific calibration standards (SSCS), are used, and the instrument's measurement of the concentrations of known analytes in the samples are measured. Empirical calibration is effective because the samples used closely match the sample matrix. SSCSs are well-prepared samples collected from the site of interest in which the concentrations of analytes have been determined by inductively coupled plasma (ICP), atomic absorption (AA), or other methods approved by the US Environmental Protection Agency (EPA). The standards should contain all the analytes of interest and interfering analytes. Manufacturers recommend that 10 to 20 calibration samples be used to generate a calibration curve.

Compton Normalization
The Compton normalization method incorporates elements of both empirical and FP calibration. A single, well-characterized standard, such as an SRM or a SSCS, is analyzed, and the data are normalized for the Compton peak. The Compton peak is produced from incoherent backscattering of X-ray radiation from the excitation source and is present in the spectrum of every sample. The intensity of the Compton peak changes as various matrices affect the way in which source radiation is scattered. For that reason, normalizing to the Compton peak can reduce problems with matrix effects that vary among samples. Compton normalization is similar to the use of internal standards in analysis for organic analytes.

Sample Preparation

Procedures for sample preparation for in situ and intrusive analysis vary considerably, since the two methods serve completely different purposes. Sample preparation for in situ analysis is fairly straightforward, while sample preparation for intrusive analysis can be fairly complicated, depending on the data quality required.

In situ or “point-and-shoot” analysis requires little sample preparation.

  • First, any unrepresentative debris, such as rocks, pebbles, leaves, vegetation, roots, and so forth, should be removed from the surface of the soil.
  • Second, the surface must be smooth, so that the probe window makes good contact with the soil surface.
  • Last, the surface of the soil should not be saturated to the point that ponded water is present.
For an "in situ prepared" sample:
  • Soil from the sampling point is collected, and all unrepresentative debris, such as rocks, pebbles, leaves, vegetation, roots, and so forth, is removed.

  • The soil is thoroughly homogenized.

  • The sample probe is placed directly on the soil for analysis, as with a true in situ sample, or the sample can be analyzed directly through a plastic bag used for homogenization.

For intrusive analysis, the sample first must be collected and then prepared for analysis in a sample cup. Some or all of the following steps are necessary, depending on the data quality needed:

  • The most important preparation step is thorough homogenization. Mixing the sample in a plastic bag works well.

  • Any large unrepresentative debris should be removed from the sample.

  • If the sample contains more than 20 percent moisture, the sample should be dried, preferably in a convection or toaster oven. Drying in a microwave oven is discouraged because doing so can increase the variability of results and arcing can occur when metal fragments are present in the sample.

  • If a high degree of precision is required, the sample should be passed through a sieve. If the sample is not wet (has a moisture content of less than 20 percent) and is not high in clay content, the sample can be sieved in the field before it is placed in a container. Otherwise, the sample is ground with a mortar and pestle and passed through a 40- or 60-mesh sieve after drying.

  • Finally, the sample is placed in a 31- or 40-mm polyethylene cup and covered with Mylar film.

Quality Control

Ensuring that the data generated by FPXRF analysis are of a known quality is vital to ensuring the usefulness of those data, regardless of their purpose. Quality control (QC) measures take several forms and can be performed in the field, during sample analysis, and after sample data have been collected. The amount and type of QC necessary will depend on the project's data quality objectives. A much higher degree of QC is necessary to produce defensible, definitive data, but analytical results from intrusive analysis have been demonstrated to compare favorably with results obtained through traditional laboratory methods, given that sample preparation has been thorough and QC adequate. By nature, results obtained in situ are of lower quality because of the lack of sample preparation, but, with the use of proper QC, in situ data can be corrected. A typical QC program would include the following measures:

  • An energy calibration check sample at least twice daily

  • An instrument blank for every 20 environmental samples
  • A method blank for every 20 samples
  • A calibration verification check sample for every 20 samples
  • A precision sample for every 20 environmental samples.
  • A confirmatory sample for every 10 environmental samples

  • Each of the measures identified above is discussed in detail below.

    Energy calibration check samples are used to test FP calibrations. A check sample consists of a pure element, such as iron, lead, or copper, and is analyzed to determine whether the characteristic X-ray lines are shifting, which would indicate drift in the detector. The check also serves as a gain check in the event that ambient temperatures are fluctuating significantly (more than 10 to 20° F). The energy calibration check should be run at a frequency consistent with the manufacturer's recommendations. Generally, the check would be performed at the beginning of each working day, after the batteries have been changed or the instrument shut off, at the end of each working day, and at any other time at which the instrument operator believes that drift is occurring during analysis.

    Two types of blanks can be used during FPXRF analysis. The first is an instrument blank, which is used to verify that there is no contamination in the spectrometer or on the probe window. The instrument blank can be silicon dioxide, a Teflon block, or a quartz block. The instrument blank should be analyzed a minimum of once daily, preferably once for every 20 samples, and should not contain any target analytes at levels higher than the MDL. The second type of blank is a method blank. The method blank is used to monitor sampling and analysis methods for laboratory-induced contaminants or interferences. The method blank can be “clean” silica sand or lithium carbonate that undergoes the same sample preparation procedures as the environmental samples. The method blank should be analyzed with the same frequency as the instrument blank and should not contain any target analytes at levels higher than the MDL.

    Precision and Accuracy

    Calibration verification check samples are used to check the accuracy of the instrument and assess the stability and consistency of the analysis of the target analytes. Accuracy is a measure of the instrument's ability to measure the “true” concentration of an element in a sample. The check sample can be an SSCS or an SRM, such as the National Institute of Standards and Technology (NIST) SRMs, that contains the target analytes, preferably at concentrations near any action levels for the site. The check sample should be run at the beginning and the end of each day or for every 20 environmental samples. The percent difference (%D) between the true value and the measured value should be less than 20 percent.

    Instrument precision refers to an instrument's ability to produce the same result for a number of measurements of the same sample. The precision of FPXRF measurements is monitored by performing several analyses of samples that contain low, medium, and high concentrations of target analytes. It is especially important to know the precision of the instrument in measuring concentrations that are similar to action levels, because precision is dependent on analyte concentrations of analytes: as the concentration increases, the precision increases. A minimum of one precision sample should be run per day by conducting from 7 to 10 replicate measurements of the sample. The precision is assessed by calculating a relative standard deviation (RSD) of the replicate measurements for the analyte. The RSD values should be less than 20 percent for most analytes, except chromium, for which the value should be less than 30 percent.

    Click to see a comparison of instrument precision.

    Click to view the percent recovery by the FPXRF instrument for a number of metals in performance evaluations and standard reference materials.

    Click to view the performance of an FPXRF instrument during the analysis of commercial performance evaluation (PE) samples. PE samples are commercially available standards containing certified concentrations of various target analytes.

    Confirmatory samples are collected from the same sample material that is analyzed on site, but are sent to an off-site laboratory for formal analysis. The results of the on-site analysis are compared with the results of the off-site analysis to determine whether they are comparable within the acceptable range. The acceptable range is determined by the analytical method, if applicable, or by the user. The purpose of a confirmatory sample is to judge the accuracy of the data obtained by analysis on site and to allow corrections, if necessary. One confirmatory sample usually is submitted for every 10 to 20 samples analyzed on site, depending on the nature of the job.

    Advantages

    Most instruments weigh less than 30 pounds and can be operated using battery power for 8 to 10 hours.

    A sample can be analyzed in less than five minutes. Throughput is a measure of the maximum rate of analysis that realistically can be obtained when using an instrument. That measure includes not only analytical time, but all sample preparation, QC, and data processing necessary to produce useable results. Throughput usually is expressed in samples per hour or samples per day. A throughput of 50 to 100 samples a day typically can be achieved for intrusive analysis, and as many as 150 samples per day can be analyzed in situ.

    Analyses of as many as 35 elements can be performed simultaneously in a single analysis.

    The sample is not destroyed during preparation or analysis; therefore, it is possible to perform replicate analyses on a sample and send the same sample for confirmatory analysis, so that comparability studies can be performed. The sample also can be archived for later use as a soil standard.

    Because no solvents or acids are used for sample extraction, no waste is generated; disposal costs therefore are eliminated.

    Operators usually can be trained in one or two days. The software is menu-driven. No data manipulation is required. Instruments are marketed for use by general scientists.

    Little or no sample preparation is required; therefore, sample throughput is enhanced and time and money are saved.

    Limitations

    Detection limits for chromium are 200 mg/kg or higher. Action levels for some elements, such as arsenic or cadmium, may be lower than the detection limits of XRF.

    Concentrations of elements in different types of soil or matrices might change, causing interferences--for example, between arsenic and lead. Site-specific calibration standards can compensate for some of those effects.

    It is difficult to obtain soil standards. One of the best sources is SRMs from NIST. Those standards cost from $200 to $500 each.

    A specific license is required to operate some FPXRF instruments. The total cost of attending a radiation safety course, obtaining the necessary paperwork, and paying the fee for the license can range from $500 to $1,000.

    The Cd-109 source should be replaced every two years. The cost of replacement is approximately $4,000 to $5,000.

    Any instrument that has a Si(Li) detector will require liquid nitrogen and a dewar (aluminum container) to hold the liquid nitrogen. This requirement adds the time and cost of obtaining and handling liquid nitrogen to cool an instrument with a Si(Li) detector before analysis can be performed.

    Cost Data

    XRF costs vary significantly. Instrument design and accessories affect instrument prices. Manufacturers listed below should be contacted directly for cost information.

    Additional Resources

    Comparing Field Portable X-Ray Fluorescence (XRF) To Laboratory Analysis Of Heavy Metals In Soil

    Niton User's Guide Version 5.0

    On-site Analysis of Metals in Liquids

    Sample Handling Strategies for Accurate Lead-in-Soil Measurements in the Field and Laboratory

    Vendor/Instrument Information

    HNU Systems, Inc. SEFA-P Analyzer
    Scitech Corporation AP Spectrum Analyzer
    TN Spectrace TN Pb Analyzer

    Verification/Evaluation Reports

    Verification of the performance of site characterization and field analytical technologies is conducted through a variety of programs. Evaluation and verification reports from EPA's Superfund Innovative Technologies Evaluation (SITE) Measuring and Monitoring Program, EPA's Environmental Technology Verification Program (ETV) program, along with links to certification statements from California EPA's (CalEPA) California Environmental Technology Certification Program, are provided below.

    Superfund Innovative Technologies Evaluation (SITE) Measuring and Monitoring Program
    The SITE Demonstration Program encourages the development and implementation of innovative treatment technologies for (1) remediation of hazardous waste sites and (2) monitoring and measurement. In the SITE Demonstration Program, the technology is field-tested on hazardous waste materials. Engineering and cost data on the innovative technologies are gathered so that potential users can assess the technology's applicability to a particular site. Data collected during the field demonstration are used to assess the performance of the technology, the potential need for pre- and post-treatment processing of the waste, applicable types of wastes and waste matrices, potential operating problems, and approximate capital and operating costs.

    See ETV reports below

    EPA's Environmental Technology Verification (ETV) Program
    EPA's Environmental Technology Verification (ETV) Program verifies the performance of innovative technologies. ETV was created to substantially accelerate the entrance of new environmental technologies into the domestic and international marketplaces. ETV verifies commercialized, private sector technologies. After the technology has been tested, the companies will receive a verification report that they can use in marketing their products. The results of the testing also are available on the Internet. The following reports from the ETV program are available for x-ray fluorescence:

    • HNU Systems SEFA-P was verified for detection and measurement of a series of inorganic analytes in soil. The primary target analytes were arsenic, barium, chromium, copper, lead, and zinc; nickel, iron, cadmium, and antimony were secondary analytes. The verification documents available consist of a verification report.
    • The Metorex X-MET 920-P and 940Field Portable X-ray Fluorescence Analyzer was verified for detection and measurement of a series of inorganic analytes in soil. The primary target analytes were arsenic, barium, chromium, copper, lead, and zinc; nickel, iron, cadmium, and antimony were secondary analytes. The verification documents available consist of a verification report.
    • The Metorex X-MET 920-MP Fluorescence Analyzer was verified for detection and measurement of a series of inorganic analytes in soil. The primary target analytes were arsenic, barium, chromium, copper, lead, and zinc; nickel, iron, cadmium, and antimony were secondary analytes. The verification documents available consist of a verification report.
    • The Niton XL Spectrum Analyzer was verified for detection and measurement of a series of inorganic analytes in soil. The primary target analytes were arsenic, barium, chromium, copper, lead, and zinc; nickel, iron, cadmium, and antimony were secondary analytes. The verification documents available consist of a verification report.
    • The Scitec MAP Spectrum Analyzer was verified for detection and measurement of a series of inorganic analytes in soil. The primary target analytes were arsenic, barium, chromium, copper, lead, and zinc; nickel, iron, cadmium, and antimony were secondary analytes. The verification documents available consist of a verification report.
    • The Spectrace TN 9000 and TN Pb Field Portable X-ray Fluorescence Analyzers were verified for detection and measurement of a series of inorganic analytes in soil. The primary target analytes were arsenic, barium, chromium, copper, lead, and zinc; nickel, iron, cadmium, and antimony were secondary analytes. The verification documents available consist of a verification report.
    California EPA's California Environmental Technology Certification Program
    CalEPA's environmental technology certification program is a voluntary program that provides participating technology developers, manufacturers, and vendors an independent, recognized third-party evaluation of the performance of new and mature environmental technologies. Developers and manufacturers define quantitative performance claims for their technologies and provide supporting documentation; CalEPA reviews that information and, when necessary, conducts additional testing to verify the claims. The technologies, equipment, and products that are proven to work as claimed are given official state certification. The certification program is voluntary and self-supporting. Companies participating in the program pay the costs of the evaluation and certification of their technologies.

    Technologies that have been certified through this program are listed below. Links are provided to the web sites that provide the Certified Environmental Technology Transfer Advisory and Certification Notice for the technologies.

    No reports available for this technology

     

    purple divider

    Disclaimer
    http://www.cluin.org/char/technologies/xrf.cfm
    Page Last Modified: May 18, 2007